Nuclear medicine

Lecture 3

  • SPECT

    • Single photon emission computed tomography

    • Want pixel size half the FWHM resolution

    • Quality control

      • Tests system performance quarterly

        • Use a jaszczak phantom filled with Tc-99m

      • Test uniformity correction map monthly

      • Test center of rotation monthly

  • Gamma camera quality control

    • Uniformity index = measures non-uniformity = lower number is better because means less non-uniformity

    • >5% = need recalibration

  • Bar phantom

    • tests spatial resolution

    • tests linearity

Nuclear medicine