XR Physics
Lecture 2
Source to image distance
Large SID used in fixed CXR because it reduces magnification
Closer source is to the target (patient) = the more the structures will be magnified
Same thing as PA vs AP
So if you have two masses, one in anterior chest and one in posterior chest they will look different sizes
Contrast
Contrast is basically how different are the gray levels
Better contrast is achieved by
Lowering kV
Lowering FOV
Post image processing
Noise = Quantum Mottle
Basically differences in gray color for same material
Decrease noise = better image = achieved by
Increase mAs = more radiation to detector
Increase kV = more radiation to detector
Post image processing
Spatial resolution
How much blur is there, ability to see small objects
Improve (increase) spatial resolution via
Decreasing focal spot size
Decrease pixel size (decreased detector element size)
Magnification has varying effect
Detector Pitch
Detector element size
This is a small detector that says how much radiation is hitting it and gives you an average
If you have one then will average all the radiation hitting it = no bueno
More detectors = can see smaller shit = better spatial resolution
Magnification
Increasing focal spot = more blur
Decreases detector blur but increases focal spot blur
Seems important - maybe look into this more
Contrast to noise ration
If you decrease FOV = less scatter = better contrast (noise unaffected) = better C/N ration
Focal spot only affects blur
Decrease tube current = more noise = worse C/N
dN = 1/sqrt of dose = Need to know this
As dose to receptor goes up = noise goes down
dN = change in noise
If you have the mAs the noise willi ncrease 40%
Leakage - small amount of radiation that escapes machine and doses the non-patient
Primary beam - what gives the dose to the patient
Scatter - any time the beam hits anything it will get redirected and go other places
Includes hitting the patient which is usually the largest source of scatter
Exposure index
Amount of radiation hitting the detector
Typical doses
150-300 for body XR (KUB)
700-1000 = Extremity = want less noise and also more radiation hitting detector because there is less soft tissue and shit
CXR
125 Kv (high)
1 mAs (low)
5 ms of time
mostly air so
AEC used
Grid used
Exposure index = 3 uGy
entrance air kerma (Kair) = 0.1 mGy
Portable CXR
80 kV
Lower Kv (Kv is the same as tube voltage) because has less scatter, do this because cannot use a grid, cannot use grid because it is bedside and pt cannot be exactly positioned correctly perpendicular and would result in artifact
1 mAs
No grid
No AEC (it is manual)
Adult KUB
80 kV
Low Kv = gives better contrast and the abdominal shit is all gray anyway so need better contrast (if fat need more)
If go lower than this the beam will not penetrate the abdomen
20 mAs
AEC used
Grid used
Extremity XR
Small focal spot = increased spatial resolution = Kv & ma are lower so can get away with it and not have too much heat
55 kV
<1 mAs
AEC optional -
Grid optional
Peds CXR
lower kV = 60 kV
Better contrast and lower radiation
lower mAs = 1
Basically same as an extremity
Exposure index is the same = still need same amount of radiation hitting detector - the dose to patient is different _because they are different size and therefore less absorbed in smaller people (i think?) but dose to the receptor is the same because need to keep same noise and shit the same
Artifacts
Lag/Ghosting
After your take a pic, the electrons stay in detector and are not cleared out, then stuff from the old image persists on the next picture taken
Usually with high Z materials (prosthesis)
Fix via put acrylic and take blank pic if needed, this usually resolves on its own
Dust on laser (CR)/Dead pixel row (DR)
Row of detector elements that stop working
Looks like a straight line for no reason
Poor calibration
AEC is seen basically
Normally the AEC is not seen because there is calibration or some shit to make them invisible
Needs to be recalibrated to fix - called flat field correction
Grid Cutoff
Need detector to be perpendicular to the grid
If it is not then a lot of the good radiation that we need to make the image is blocked by the grid
Correct by making beam perpendicular to grid
Can lines and if radiation does not hit detector or is blocked by detector by the grid will be white (as if it were bone)
Gridlines
Multiple horizontal lines - looks like beat bord
Should not see normally because one of the follow
Grid is shaken so cannot see
Some other shit
If you see gridlines —> it stopped shaking or post image processing failed
Correct via removing grid (or fix the reason it failed, motor to shake not working)
Electromagnetic interference
Only on digital detectors
looks like bunch of lines similar to gridlines
Usually because there is a device that interferes with the XR machine by giving out electromagnetic radiation and interefers with sending shit to the computer (such as pain pump or some shit)
Effective dose
CXR = 0.02 mSv = should know this number
KUB = 0.7 mSv
d
Sources: