Fluoro Physics

Fluoroscopy

  • Basically many shitty radiographs with smaller radiation dose played together

  • ABC

    • automatic brightness control, with the detector, under the grid

  • Differences from radiography

    • smaller focal spot (0.3-1.2 mm)

      • much less heating because use in pulse mode rather than a continuous beam of radiation

    • Pulsed beam (7.5-15 pps or fps)

    • Low tube current

      • <1 to 10 mA (low mA)

      • This is why the image is shitty

      • About 1%% of the dose as a normal XR

  • Image intensifier (same as detector for fluoro)

    • CsI

    • Takes XR in —> hits CsI —> puts out light —> hits a cathode —> turns into low energy electrons —> pass through a current which makes electrons move faster through the intensifier —-> hits output phospher —> electrons converted back to light —> light goes to a camera

      • cannot go from original light because it is too faint and would have to give deadly amounts of radiation to get same dose as if you were to just amplify it as here

      • How is light amplified

        • flux gain

          • moves along electric field to make it gain energy

          • multiplies it about 100x

          • Flux gain ~50

          • Basically if you were to point a flashlight at a wall it would vaguely light up the wall but if you start walking closer to the wall it will have a smaller area that is brighter even though the flashlight is still putting out the same amount of power

    • Takes small amount of light and amplifies it about 5000x

  • Display monitor

    • Basically image has too much data to put on screen at once so came up with ways to get around that

    • <30 Hz interlaced

      • Put image up and then update every other line

      • lowest frame rate that avoids flicker

    • <60 Hz progressive

  • Collimation

    • Lower dose to patient

    • Less scatter

    • Improves image quality

  • Temporal filter (recursive frame average)

    • Taking a few frames that were already taken and averaging them together

    • Decreases noise

    • Improves signal to noise ratio by the square root of the number of frames

      • If recursive filtering averages 4 frames the signal to noise ratio increases by 2X

    • Worsens temporal resolution (worsened motion artifact)

    • Decreasing recursive filtering improves temporal resolution and vice versa

      • Want high temporal resolution = means less motion blur

  • Monitor

    • refreshes images every 33 ms

    • at 30 image will be shown once on screen before next refresh

    • at 15 pps then monitor will refresh every 2 images you take

      • each acquired frame shown twice on the screen

        • if it was 7.5 then 4x and so on

  • Contrast media

    • want 75 kV when able

  • Barium meal

    • 110 kV (higher than iodine level), allows you to see through bowel a bit, not just whited out

      • This is above the k-edge on purpose

  • Acquisition (exposure) Photospot

    • lower noise than spot frame image

    • larger focal spot and smaller pixel size vs spot

    • 10 spot images = 1 min fluoro

  • Cine

    • 8-15x fluoro dose

    • larger focal spot

Fluoroscopy Lecture 5 - Fluoro lecture 1 - Thislecture discusses analog systems

  • Each dose is about 1/100th the receptor dose of a regular radiograph

    • Comes with tradeoff of increased noise

  • ABC = Automatic brightness control

  • Grid is used

  • Differences compared to radiography

    • Can use smaller focal spots

    • Beam is pulsed (not continuous I guess) = pps or fps, same shit

      • 7.5-15 fps is typical

    • Tube current = <1-10 mA (low vs XR)

  • Image intensifier

    • CsI scintillator

    • This basically takes in the XR and small amount of light and amplifies the light by 5000x and makes it bright so we can see shit, otherwise you’d have to give crazy radiation to see stuff because its so dark

    • Takes in XR and makes it light basically

      • Light hits the photocathode

        • We basically bottle neck the photons to a small area and it hits a photophospher which give

      • Minification gain = bottle neck to a small area so the light is more focused

      • Flux gain = the gradient we make to force the photons (or whatever particle) to float to the bottlenecked area

  • Once the light comes out of the output phospher (the thing which absorbs photons and puts out the light)

    • We now collect that light and put it to the monitor and you have an image

  • ABC

    • Monitors how much light comes out from the output phospher

    • It knows how much light we need to see shit

    • If the image is too dark or bright it will adjust by playing with mA & kV so that the brightness is good

    • Similar to the AEC, but will how dark or bright it is

  • Display monitor

    • Interlaced vs progressive monitors

    • Progressive = shows 60 fps with a new image every 1/60th of a second (60 Hz) = better

    • Interlaced = new image every 1/30th of a second, basically does every line to trick your eye rather than the progressive which does every line - get weird overlap artifact

  • Last image hold

    • Shows the last frame you took, used to be that you could only see it while you are on the pedal dosing the pt

    • Loop hold is the same thing but for the cine

  • Collimation

    • Virtual collimation is when it shows you where the borders would be on last image hold so you dont have to fluoro to see what you are collimating

    • Improves image quality

      • Decreases scatter

      • Increases

      • Post image processing is done on a smaller area so is more efficient

    • Temporal filter (Recursive frame average)

      • Averages the frame you are currently fluoring + a few priors averaged together

        • This decreases noise = better signal to noise ratio

      • If you decreases temporal filter = faster to get image

        • Because if you have less images then it is easier to do? idk

      • If 4 frames averaged the SNR ratio increases 2X idk why some math shit

  • Refresh shit

    • 30 Hz is basically 30 pps

  • Fluoro systems

  • Spacer cone = makes you have the pateint a set distance away from the focal spot because dose increases significantly there and dont want too much focally

  • 75 kV is used typically

  • 110 kV for barium shit

    • Above the k edge so that the barium itself is not too bright and then cannot see shit

  • Acquisitions (exposures)

    • Larger focal spot

    • Smaller pixel size

    • 1 min of fluoror is ~ 10 exposures in terms of dose

    • Less noise vs fluoro (similar to radiograph)

  • Cine

    • 8-15x fluoro dose

    • large focal spot

  • w

  • w

  • w

  • w

Fluoro Physics