Fluoro Physics
Fluoroscopy
Basically many shitty radiographs with smaller radiation dose played together
ABC
automatic brightness control, with the detector, under the grid
Differences from radiography
smaller focal spot (0.3-1.2 mm)
much less heating because use in pulse mode rather than a continuous beam of radiation
Pulsed beam (7.5-15 pps or fps)
Low tube current
<1 to 10 mA (low mA)
This is why the image is shitty
About 1%% of the dose as a normal XR
Image intensifier (same as detector for fluoro)
CsI
Takes XR in —> hits CsI —> puts out light —> hits a cathode —> turns into low energy electrons —> pass through a current which makes electrons move faster through the intensifier —-> hits output phospher —> electrons converted back to light —> light goes to a camera
cannot go from original light because it is too faint and would have to give deadly amounts of radiation to get same dose as if you were to just amplify it as here
How is light amplified
flux gain
moves along electric field to make it gain energy
multiplies it about 100x
Flux gain ~50
Basically if you were to point a flashlight at a wall it would vaguely light up the wall but if you start walking closer to the wall it will have a smaller area that is brighter even though the flashlight is still putting out the same amount of power
Takes small amount of light and amplifies it about 5000x
Display monitor
Basically image has too much data to put on screen at once so came up with ways to get around that
<30 Hz interlaced
Put image up and then update every other line
lowest frame rate that avoids flicker
<60 Hz progressive
Collimation
Lower dose to patient
Less scatter
Improves image quality
Temporal filter (recursive frame average)
Taking a few frames that were already taken and averaging them together
Decreases noise
Improves signal to noise ratio by the square root of the number of frames
If recursive filtering averages 4 frames the signal to noise ratio increases by 2X
Worsens temporal resolution (worsened motion artifact)
Decreasing recursive filtering improves temporal resolution and vice versa
Want high temporal resolution = means less motion blur
Monitor
refreshes images every 33 ms
at 30 image will be shown once on screen before next refresh
at 15 pps then monitor will refresh every 2 images you take
each acquired frame shown twice on the screen
if it was 7.5 then 4x and so on
Contrast media
want 75 kV when able
Barium meal
110 kV (higher than iodine level), allows you to see through bowel a bit, not just whited out
This is above the k-edge on purpose
Acquisition (exposure) Photospot
lower noise than spot frame image
larger focal spot and smaller pixel size vs spot
10 spot images = 1 min fluoro
Cine
8-15x fluoro dose
larger focal spot
Fluoro Physics